FTB-5240S/S-P/BP

Product-Specific Commands—Description

:UNIT[1..n]:SPECtrum?

Description

This query returns the selected default unit for
commands which program spectrum.

At *RST, default spectrum unit is set to M (meter).

Syntax

:UNIT[1..n]:SPECtrum?

Parameter(s)

None

Response Syntax

<Unit>

Response(s)

Unit:
The response data syntax for <Unit> is defined as
a <CHARACTER RESPONSE DATA> element.

The response corresponds to the selected default
spectrum unit.

M: meter unit is selected.
HZ: hertz unit is selected.

Example(s)

UNIT:SPEC HZ
UNIT:SPEC? Returns HZ

See Also

:UNIT[1..n]:POWer
:UNIT[1..n]:RATio
:UNIT[1..n]:SPECtrum