Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 558

Advertising
background image

Справка по командам SCPI

552

FTB-5240S/S-P/BP

Product-Specific Commands—Description

Response(s)

Resolution:
The response data syntax for <Resolution> is
defined as a <NR3 NUMERIC RESPONSE
DATA> element.

The <Resolution> response corresponds to either
the current or the MINimum/MAXimum resolution
bandwidth value for OSNR calculation.

Example(s)

CALC:WDM:OSNR:BAND:RES:AUTO OFF
CALC:WDM:OSNR:BAND:RES 0.065 NM
CALC:WDM:OSNR:BAND:RES? Returns
6.500000E-011

See Also

:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]:AUTO
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]

:CALCulate[1..n][:WDM]:OSNR:

BANDwidth|BWIDth[:RESolution]?