571

Product-Specific Commands—Description

:INITiate:CONTinuous?

Description

This query indicates if the trigger system is
continuously initiated or not.

At *RST, this value is set to off (disabled).

Syntax

:INITiate:CONTinuous?

Parameter(s)

None

Response Syntax

<Continuous>

Response(s)

Continuous:
The response data syntax for <Continuous> is
defined as a <NR1 NUMERIC RESPONSE
DATA> element.

The <Continuous> response corresponds to the
state of the trigger system continuous cycle.

0: continuous cycle is disabled.
1: continuous cycle is enabled.

Example(s)

INIT:CONT ON
INIT:CONT? Returns 1 (trigger system
continuously initiated)

Notes

An acquisition may still be in progress even if
INIT:CONT? returns 0. The command
STAT:OPER:BIT8:COND? shall be used to test
acquisition completion.

See Also

:INITiate[:IMMediate]
:INITiate:CONTinuous
:STATus?
:STATus:OPERation:BIT<8|9>:CONDition?