Справка по командам SCPI

634

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:TRIGger[1..n][:SEQuence]:SOURce

Description

This command selects the source for the trigger
system event detector.

At *RST, the source is set to IMMediate.

Syntax

:TRIGger[1..n][:SEQuence]:SOURce<wsp>IMMe
diate|TIMer

Parameter(s)

Source:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
IMMediate|TIMer.

The parameter corresponds to the newly selected
trigger event source.

IMMediate: No waiting for an event occurs.
TIMer: The source signal comes from a periodic
timer.

Example(s)

TRIG:SEQ:SOUR IMM
TRIG:SEQ:SOUR? Returns IMMEDIATE

Notes

The TIMer trigger event source is valid for queries
only. It is used internally during drift acquisition.

See Also

:INITiate[:IMMediate]
:INITiate:CONTinuous
:TRIGger[1..n][:SEQuence]:SOURce?