Справка по командам SCPI

OSA

577

Product-Specific Commands—Description

Parameter(s)

The <ColumnName> contains a comma
separated list of the name of the column to
include in the table. The column order is
preserved. Unrecognized column definition will
produce empty result. Duplicates are allowed.
The possible entries in this list are any of the
following elements: BAND1:FREQ,
BAND1:RLEV, BAND1:WAV, BAND2:FREQ,
BAND2:RLEV, BAND2:WAV, BWID1:FREQ,
BWID1:RLEV, BWID1:WAV, BWID2:FREQ,
BWID2:RLEV, BWID2:WAV, CENT:FREQ,
CENT:WAV, CMAS:FREQ, CMAS:WAV,
CPEA:FREQ, CPEA:WAV, ENBW, NAME, NOIS,
NOIS:TYPE, OSNR, SIGP, SIGP:TYPE,
STAT:QUES:BIT9:COND,
STAT:QUES:BIT10:COND,
STAT:QUES:BIT11:COND, WIDT:FREQ or
WIDT:WAV. Consult the
:CALCulate:WDM:DATA:CHANnel command tree
to get a description of the return value for the
previous elements. Only short form is accepted.

Example(s)

<Do measurement>
MEM:TABL:SEL "WDM:CHANNEL"
MEM:TABL:DEF "NAME,CMAS:WAV"
MEM:TABL:POIN? "WDM:CHANNEL" returns 2
MEM:TABL:DATA? "WDM:CHANNEL" returns
#248"C_001,1.55236113E-006","C_002,1.55672
7357E-006"

Notes

The only valid table name is "WDM:Channel".

See Also

:MEMory:TABLe:DATA?
:MEMory:TABLe:DEFine?
:MEMory:TABLe:SELect

:MEMory:TABLe:DEFine