Calculate[1..n][:wdm]:data: signalpower:flatness – Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 552

Advertising
background image

Справка по командам SCPI

546

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:DATA:

SIGnalPower:FLATness?

Description

This query returns computed WDM analysis
global result for signal power flatness.

At *RST, this value is not available.

Syntax

:CALCulate[1..n][:WDM]:DATA:SIGnalPower:FLA
Tness?

Parameter(s)

None

Response Syntax

<Flatness>

Response(s)

Flatness:
The response data syntax for <Flatness> is
defined as a <NR3 NUMERIC RESPONSE
DATA> element.

The <Flatness> response corresponds to the
computed signal power flatness.

Example(s)

<Do measurement>
UNIT:RAT DB
CALC:DATA:SIGP:FLAT? Returns
3.118000E+001

Notes

Special NAN (not a number) value
-2251799813685248 is returned if analysis result
could not be computed.

See Also

:CALCulate[1..n][:WDM]:DATA:OSNR:FLATness
?
:CALCulate[1..n][:WDM]:DATA:OSNR:MEAN?
:CALCulate[1..n][:WDM]:DATA:SIGnalPower:ME
AN?
:CALCulate[1..n][:WDM]:DATA:TPOWer?

Advertising