Справка по командам SCPI

486

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:

NOISe:WIDTh[:WAVelength]?

Description

This query returns the wavelength width of the
noise measurement region of the selected WDM
analysis channel.

At *RST, this value is not available.
At CALCulate[1..n][:WDM]:CHANnel[:DEFine],
width of the noise measurement region is set to
100.0 GHz.

Syntax

:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh
[:WAVelength]?[<wsp>MAXimum|MINimum|DEF
ault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.

MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.

Response Syntax

<Width>