Product-Specific Commands—Description

:SENSe[1..n]:AVERage:STATe?

Description

This query indicates if acquired trace averaging
has been enabled or not.

At *RST, averaging is set to off (disabled).

Syntax

:SENSe[1..n]:AVERage:STATe?

Parameter(s)

None

Response Syntax

<State>

Response(s)

State:
The response data syntax for <State> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.

The <State> response corresponds to the
activation state of trace averaging.

0: trace averaging is disabled.
1: trace averaging is enabled.

Example(s)

SENS:AVER:STAT ON
SENS:AVER:STAT? Returns 1 (averaging is
enabled)

See Also

:SENSe[1..n]:AVERage:COUNt
:SENSe[1..n]:AVERage:TYPE