Справка по командам SCPI

590

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:SENSe[1..n]:AVERage:TYPE

Description

This command selects the acquired trace
averaging type.

At *RST, averaging is set to SCALar.

Syntax

:SENSe[1..n]:AVERage:TYPE<wsp>SCALar|Pol
arizationMinMaxHold

Parameter(s)

Type:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
SCALar|PolarizationMinMaxHold.

The parameter corresponds to the newly selected
trace averaging type.

SCALar: selects scalar averaging type.
PolarizationMinMaxHold: selects averaging type
for InBand noise measurement.

Example(s)

SENS:AVER:TYPE SCAL
SENS:AVER:TYPE? Returns SCALAR