Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 557

Advertising
background image

Справка по командам SCPI

OSA

551

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:OSNR:

BANDwidth|BWIDth[:RESolution]?

Description

This query returns a value indicating either the
current or the minimum/maximum resolution
bandwidth value for WDM analysis OSNR
calculation.

At *RST, this value is set to 0.100 nm.

Syntax

:CALCulate[1..n][:WDM]:OSNR:BANDwidth|BWI
Dth[:RESolution]?[<wsp>MAXimum|MINimum|DE
Fault]

Parameter(s)

Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.

MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.

Response Syntax

<Resolution>

Advertising