Sense[1..n]:average:count – Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 590

Advertising
background image

Справка по командам SCPI

584

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:SENSe[1..n]:AVERage:COUNt

Description

This command sets the number of acquired
traces to combine for averaging to a specific
value.

At *RST, averaging count is set to 8.

Syntax

:SENSe[1..n]:AVERage:COUNt<wsp><Count>|M
AXimum|MINimum|DEFault

Parameter(s)

Count:
The program data syntax for <Count> is defined
as a <numeric_value> element. The <Count>
special forms MINimum, MAXimum and DEFault
are accepted on input.

MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <Count> parameter.

The <Count> parameter corresponds to a valid
averaging count value.

The SENSe[1..n]:AVERage:COUNt? MIN and
SENSe[1..n]:AVERage:COUNt? MAX queries
can be used to determine valid count range.