Product-specific commands—description, Abort – Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 401

Advertising
background image

Справка по командам SCPI

OSA

395

Product-Specific Commands—Description

Product-Specific Commands—Description

:ABORt

Description

This command resets the trigger system and
places all trigger sequences in the IDLE state.
Any trace acquisition that is in progress is aborted
as quickly as possible. The command is not
completed until the trigger sequence is in the
IDLE state.

This command is an event and has no associated
*RST condition or query form.

Syntax

:ABORt

Parameter(s)

None

Example(s)

ABOR

Notes

A call to ABORt only returns once acquisition is
completely stopped and instrument is ready for
new commands. For this reason, execution of this
command may take a few seconds.

For a continuously initiated acquisition
(INIT:CONT ON), calling ABORt will automatically
set it to OFF.

See Also

:INITiate[:IMMediate]
:INITiate:CONTinuous
:STATus
:STATus:OPERation:BIT<8|9>:CONDition?

Advertising