Справка по командам SCPI

OSA

569

Product-Specific Commands—Description

With <Continuous> set to OFF, the trigger system
remain in idle state until <Continuous> is set to
ON or INITiate:IMMediate command is received.
With <Continuous> set to ON, the trigger system
leave idle state and continue cycling until
<Continuous> set to OFF or ABORt command is
received.

When <Continuous> is set to OFF, the current
trigger cycle is completed before returning to idle
state: current acquisition continue until finished.

Example(s)

CALC:WDM:CHAN:DEL:ALL
CALC:WDM:CHAN:AUTO ON
CALC:WDM:CHAN:AUTO:NOIS:AUTO OFF
CALC:WDM:CHAN:AUTO:NOIS:TYPE IEC
CALC:WDM:STATe ON
TRACe:FEED:CONTrol "TRC1", ALW
SENS:AVER:STAT OFF

:INITiate:CONTinuous