Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 402

Advertising
background image

Справка по командам SCPI

396

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:

BANDwidth[1|2]|BWIDth[1|2]:

RelativeLEVel

Description

This command sets WDM analysis bandwidth
position for all channels to a specific value.
Bandwidth position value is the power level
relative to peak maximum where signal bandwidth
of a channel is computed.

At *RST, this value is set to 3.0 dB for bandwidth1
and 20.0 dB for bandwidth2.

Syntax

:CALCulate[1..n][:WDM]:BANDwidth[1|2]|BWIDth[
1|2]:RelativeLEVel<wsp><PowerLevel[<wsp>DB|
W/W|PCT]>|MAXimum|MINimum|DEFault

Parameter(s)

PowerLevel:
The program data syntax for <PowerLevel> is
defined as a <numeric_value> element followed
by an optional <SUFFIX PROGRAM DATA>
element. The allowed <SUFFIX PROGRAM
DATA> elements are: DB|W/W|PCT. The
<PowerLevel> special forms MINimum,
MAXimum and DEFault are accepted on input.

MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.

Advertising