Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)
Страница 486
Advertising
Справка по командам SCPI
480
FTB-5240S/S-P/BP
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Width> parameter.
The <Width> parameter corresponds to a valid
width in hertz for the noise measurement region.
The
CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh:
FREQuency? MIN and
CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh:
FREQuency? MAX queries can be used to
determine valid width values.
Example(s)
CALC:WDM:CHAN:DEF "ITU_22",192.1750 THZ
CALC:WDM:CHAN:SEL "ITU_22"
CALC:WDM:CHAN:NOIS:TYPE POLY5
CALC:WDM:CHAN:NOIS:WIDTh:FREQ 75.0
GHZ
CALC:WDM:CHAN:NOIS:WIDTh:FREQ?
Returns 7.500000E+010
:CALCulate[1..n][:WDM]:CHANnel:
NOISe:WIDTh:FREQuency
Advertising