Справка по командам SCPI

480

FTB-5240S/S-P/BP

Product-Specific Commands—Description

DEFault allows the instrument to select a value
for the <Width> parameter.

The <Width> parameter corresponds to a valid
width in hertz for the noise measurement region.

The
CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh:
FREQuency? MIN and
CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh:
FREQuency? MAX queries can be used to
determine valid width values.

Example(s)

CALC:WDM:CHAN:DEF "ITU_22",192.1750 THZ
CALC:WDM:CHAN:SEL "ITU_22"
CALC:WDM:CHAN:NOIS:TYPE POLY5
CALC:WDM:CHAN:NOIS:WIDTh:FREQ 75.0
GHZ
CALC:WDM:CHAN:NOIS:WIDTh:FREQ?
Returns 7.500000E+010

:CALCulate[1..n][:WDM]:CHANnel: