Справка по командам SCPI

556

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:STATe

Description

This command controls activation of WDM
analysis.

At *RST, WDM analysis is set to on (enabled).

Syntax

:CALCulate[1..n][:WDM]:STATe<wsp><Auto>

Parameter(s)

Auto:
The program data syntax for <Auto> is defined as
a <Boolean Program Data> element. The <Auto>
special forms ON and OFF are accepted on input
for increased readability. ON corresponds to 1
and OFF corresponds to 0.

The <Auto> parameter corresponds to the new
state of WDM analysis.

0 or OFF: WDM analysis is disabled.
1 or ON: WDM analysis is enabled.

Example(s)

CALC:WDM:STAT ON
CALC:WDM:STAT? Returns 1 (WDM analysis
enabled)

Notes

WDM analysis cannot be disabled: OFF (0) value
is valid for queries only.

It is possible to query acquired trace data only if
active measurement analysis mode is WDM.

See Also

:CALCulate[1..n][:WDM]:STATe?
:INITiate[:IMMediate]
:INITiate:CONTinuous
:TRACe:FEED:CONTrol