Справка по командам SCPI

OSA

397

Product-Specific Commands—Description

DEFault allows the instrument to select a value
for the <PowerLevel> parameter.

The <PowerLevel> parameter corresponds to a
valid bandwidth position value.

The CALCulate[1..n]:BANDwidth? MIN and
CALCulate[1..n]:BANDwidth? MAX queries can
be used to determine valid bandwidth position
range.

Example(s)

UNIT:RAT DB
CALC:BWID2:RLEV 10.55 DB
CALC:BWID2:RLEV? Returns: 1.055000E+001
CALC:WDM:BAND2:RLEV DEF
CALC:WDM:BAND2:RLEV? Returns:
2.000000E+001

Notes

Bandwidth1 position cannot be changed: it is
always set at 3.0 dB.

See Also

:CALCulate[1..n][:WDM]:BWIDth[1|2]|BANDwidth[
1|2]:RelativeLEVel?
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]:AUTO

:CALCulate[1..n][:WDM]:THReshold

:CALCulate[1..n][:WDM]:

BANDwidth[1|2]|BWIDth[1|2]:

RelativeLEVel