Справка по командам SCPI

420

FTB-5240S/S-P/BP

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:AUTO:

NOISe:WIDTh:[WAVelength]

Description

This command sets the wavelength width of the
noise measurement region of the WDM analysis
default channel.

At *RST, the width of the default channel noise
measurement region is set to 100.0 GHz.

Syntax

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh:[WAVelength]<wsp><Width[<wsp>M]>|M
AXimum|MINimum|DEFault

Parameter(s)

Width:
The program data syntax for <Width> is defined
as a <numeric_value> element followed by an
optional <SUFFIX PROGRAM DATA> element.
The allowed <SUFFIX PROGRAM DATA>
element is M. The <Width> special forms
MINimum, MAXimum and DEFault are accepted
on input.

MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.