Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 523

Advertising
background image

Справка по командам SCPI

OSA

517

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:DATA:CHANnel:

CenterMASs[:WAVelength]?

Description

This query returns computed WDM analysis result
for center of mass wavelength of the selected
channel.

At *RST, this value is not available.

Syntax

:CALCulate[1..n][:WDM]:DATA:CHANnel:Center
MASs[:WAVelength]?

Parameter(s)

None

Response Syntax

<Center>

Response(s)

Center:
The response data syntax for <Center> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.

The <Center> response corresponds to the
computed center of mass wavelength in meter.

Example(s)

CALC:WDM:CHAN:DEL:ALL
CALC:WDM:CHAN:DEF "C_2", 1287.000 NM
<Do measurement>
CALC:WDM:DATA:CHAN:SEL "C_2"
CALC:WDM:DATA:CHAN:CMAS:WAV? Returns
1.286971E-006