Status:operation:bit<8|9>: condition – Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 626

Advertising
background image

Справка по командам SCPI

620

FTB-5240S/S-P/BP

Product-Specific Commands—Description

The <Condition> response represents the current
operation condition of the instrument. The
meaning of the response depends on the value
returned for bit <n>.

bit <8>: When the returned value is 1, the
instrument is performing a measurement (trigger
system INITiated).

bit <9>: When the returned value is 1, the
instrument is performing an offset nulling and/or a
wavelength referencing
(CALibration:ZERO:AUTO?).

Example(s)

STAT? Must return READY
CAL:ZERO:AUTO ONCE
STAT:OPER:BIT9:COND? Keep resending this
query as long as the operation is not complete
(returned value is not 0).

See Also

:CALibration:ZERO:AUTO?
:INITiate[:IMMediate]
:INITiate:CONTinuous?
:STATus?

:STATus:OPERation:BIT<8|9>:

CONDition?

Advertising