Calculate[1..n][:wdm]:data:channel: signalpower – Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 546

Advertising
background image

Справка по командам SCPI

540

FTB-5240S/S-P/BP

Product-Specific Commands—Description

Response(s)

Signal:
The response data syntax for <Signal> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.

The <Signal> response corresponds to the
computed signal power level.

Example(s)

CALC:CHAN:DEL:ALL
CALC:CHAN:DEF "ITU_019", 229.7860 THZ
CALC:CHAN:SEL "ITU_019"
CALC:CHAN:SIGP:TYPE TPOW
<Do measurement>
CALC:DATA:CHAN:SEL "ITU_019"
UNIT:POW DBM
CALC:DATA:CHAN:SIGP? Returns
-3.430000E+000

See Also

:CALCulate[1..n][:WDM]:CHANnel:AUTO:SIGnal
Power:TYPE
:CALCulate[1..n][:WDM]:CHANnel:SIGnalPower:
TYPE
:CALCulate[1..n][:WDM]:DATA:CHANnel:SIGnal
Power:TYPE?
:CALCulate[1..n][:WDM]:DATA:CHANnel:SELect
:CALCulate[1..n][:WDM]:DATA:CHANnel::STATu
s:QUEStionable:BIT<9|10|11>:CONDition?

:CALCulate[1..n][:WDM]:DATA:CHANnel:

SIGnalPower?