Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 521

Advertising
background image

Справка по командам SCPI

OSA

515

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:DATA:CHANnel:

CenterMASs:FREQuency?

Description

This query returns computed WDM analysis result
for center of mass frequency of the selected
channel.

At *RST, this value is not available.

Syntax

:CALCulate[1..n][:WDM]:DATA:CHANnel:Center
MASs:FREQuency?

Parameter(s)

None

Response Syntax

<Center>

Response(s)

Center:
The response data syntax for <Center> is defined
as a <NR3 NUMERIC RESPONSE DATA>
element.

The <Center> response corresponds to the
computed center of mass frequency in hertz.

Example(s)

CALC:CHAN:DEL:ALL
CALC:CHAN:DEF "ITU_14", 201.9873 THZ
<Do measurement>
CALC:DATA:CHAN:SEL "ITU_14"
CALC:DATA:CHAN:CMAS:FREQ? Returns
2.020066E+014

Advertising