Calculate[1..n][:wdm]:channel: signalpower:type – Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 503

Advertising
background image

Справка по командам SCPI

OSA

497

Product-Specific Commands—Description

:CALCulate[1..n][:WDM]:CHANnel:

SIGnalPower:TYPE

Description

This command selects signal power
measurement type for WDM analysis of the
selected channel.

At *RST, this value is not available.
At CALCulate[1..n][:WDM]:CHANnel[:DEFine],
this value is set to IPOWer (integrated power).

Syntax

:CALCulate[1..n][:WDM]:CHANnel:SIGnalPower:
TYPE<wsp>IPOWer|PPOWer|TPOWer

Parameter(s)

Type:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
IPOWer|PPOWer|TPOWer.

The parameter corresponds to the newly selected
signal power type.

IPOWer: selects integrated signal power type.
PPOWer: selects peak signal power type.
TPOWer: selects channel total power type.