Product-Specific Commands—Description

:SENSe[1..n]:AVERage:STATe

Description

This command controls activation of acquired
trace averaging.

At *RST, averaging is set to off (disabled).

Syntax

:SENSe[1..n]:AVERage:STATe<wsp><State>

Parameter(s)

State:
The program data syntax for <State> is defined
as a <Boolean Program Data> element. The
<State> special forms ON and OFF are accepted
on input for increased readability. ON
corresponds to 1 and OFF corresponds to 0.

The <State> parameter corresponds to the new
state of trace averaging.

0 or OFF: disable averaging.
1 or ON: enable averaging.

Example(s)

SENS:AVER:STAT OFF
SENS:AVER:STAT? Returns 0 (averaging is
disabled)

Notes

Trace averaging is not supported by the trigger
system when continuously initiated (INIT:CONT
ON).

See Also

:INITiate[:IMMediate]
:INITiate:CONTinuous
:SENSe[1..n]:AVERage:COUNt
:SENSe[1..n]:AVERage:TYPE
:SENSe[1..n]:AVERage:STATe?