Справка по командам SCPI

OSA

417

Product-Specific Commands—Description

DEFault allows the instrument to select a value
for the <Width> parameter.

The <Width> parameter corresponds to a valid
width in hertz for the noise measurement region.

The
CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh:FREQuency? MIN and
CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh:FREQuency? MAX queries can be used
to determine valid width values.

Example(s)

CALC:WDM:CHAN:AUTO:NOIS:TYPE POLY5
CALC:WDM:CHAN:AUTO:NOIS:WIDTh:FREQ
100.0 GHZ
CALC:WDM:CHAN:AUTO:NOIS:WIDTh:FREQ?
Returns 1.000000E+011

Notes

Custom width for noise measurement region is
applied only if selected noise type is
POLYnomial5.

See Also

:CALCulate[1..n][:WDM]:CHANnel:AUTO
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh[:WAVelength]
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh:FREQuency?

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:FREQuency
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
TYPE
:CALCulate[1..n][:WDM]:CHANnel:NOISe:WIDTh
:FREQuency

:CALCulate[1..n][:WDM]:CHANnel:AUTO:

NOISe:WIDTh:FREQuency