Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 415

Advertising
background image

Справка по командам SCPI

OSA

409

Product-Specific Commands—Description

DEFault allows the instrument to select a value
for the <Distance> parameter.

The <Distance> parameter corresponds to a valid
distance in hertz from peak to center of noise
region.

The
CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:FREQuency? MIN and
CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:FREQuency? MAX queries can be
used to determine valid distance values.

Example(s)

CALC:WDM:CHAN:AUTO:NOIS:TYPE POLY5
CALC:WDM:CHAN:AUTO:NOIS:DIST:FREQ
100.0 GHZ
CALC:WDM:CHAN:AUTO:NOIS:DIST:FREQ?
Returns 1.000000E+011

Notes

Custom noise measurement distance is applied
only if selected noise type is POLYnomial5.

See Also

:CALCulate[1..n][:WDM]:CHANnel:AUTO
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance[:WAVelength]
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:FREQuency?

:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh:FREQuency
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
TYPE
:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTa
nce:FREQuency

:CALCulate[1..n][:WDM]:CHANnel:AUTO:

NOISe:DISTance:FREQuency