Инструкция по эксплуатации EXFO FTB -5240S/5240BP Optical Spectrum Analyzer for FTB 500 (NS-1434)

Страница 556

Advertising
background image

Справка по командам SCPI

550

FTB-5240S/S-P/BP

Product-Specific Commands—Description

DEFault allows the instrument to select a value
for the <Resolution> parameter.

The <Resolution> parameter corresponds to the
custom resolution bandwidth in meter.

The
CALCulate[1..n][:WDM]:OSNR:BANDwidth[RESo
lution]? MIN and
CALCulate[1..n][:WDM]:OSNR:BANDwidth[RESo
lution]? MAX queries can be used to determine
valid resolution bandwidth range.

Example(s)

CALC:WDM:OSNR:BAND:RES:AUTO OFF
CALC:WDM:OSNR:BAND:RES 0.065 NM
CALC:WDM:OSNR:BAND:RES? Returns
6.500000E-011

See Also

:CALCulate[1..n][:WDM]:BWIDth[1|2]|BANDwidth[
1|2]:RelativeLEVel
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]?
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDwi
dth[:RESolution]:AUTO
:CALCulate[1..n][:WDM]:THReshold

:CALCulate[1..n][:WDM]:OSNR:

BANDwidth|BWIDth[:RESolution]

Advertising