Product-Specific Commands—Description

:MEMory:TABLe:SELect

Description

This command selects the table to define.

At *RST, there is no selection: a single null string
is returned.

Syntax

:MEMory:TABLe:SELect<wsp><TableName>

Parameter(s)

TableName:
The program data syntax for <TableName> is
defined as a <STRING PROGRAM DATA>
element.

The <TableName> parameter corresponds to the
name of the table to select.

Example(s)

<Do measurement>
MEM:TABL:SEL "WDM:CHANNEL"
MEM:TABL:DEF "NAME,CMAS:WAV"
MEM:TABL:POIN? "WDM:CHANNEL" returns 2
MEM:TABL:DATA? "WDM:CHANNEL" returns
#248"C_001,1.55236113E-006","C_002,1.55672
7357E-006"

Notes

The only valid table name is "WDM:Channel".

See Also

:MEMory:TABLe:DEFine
:MEMory:TABLe:DEFine?
:MEMory:TABLe:SELect?