Fluke 434-II/435-II/437-II

Руководство пользователя

C-2

Security Summary:

Ad 1. Memory contents erased at power-off. No user access.
Ad 2. Memory contents erased at power-off. No user access.
Ad 3. Flash memory: contents stays available at power off and disconnection of the Li-

ion accumulator (can be loaded/exchanged with dedicated PC software that is

exclusively available in manufacturing and Fluke service). Note: the calibration

data is generated when the analyzer is sent through its calibration process and are

fundamental to the meter operation.

To erase Analyzer settings that differ from Factory Default, do the following key

operations: SETUP, function key F1 – USER PREF, F1 – FACTORY

DEFAULTS, F5 – YES (confirm menu).

Ad 4. Memory contents erased at power-off. No user access.
Ad 5. Memory contents erased at power-off. No user access.
Ad 6. There are 2 ways of removing measurement data from the Analyzer:

1 - The SD Card is located in the battery-compartment at the rear of the

Analyzer. Open the compartment with a small screwdriver. Push the SD Card in

the direction of the arrow and take it out of the Analyzer. All measurement data

now has been removed from the Analyzer. Avoid touching the contacts of the

Card with you hands. When reinstalling the Card take careful notice of the

indication in the battery compartment.

2 - All measurement data at the SD memory card is erased by formatting it. The

format action occurs via a confirm menu. Do the following key operations with

the SD Card installed in the Analyzer: SETUP, function key F1 – USER PREF,

F4 – FORMAT SD CARD, F5 – YES.