Продолжение плана меню hart ® vegason 51 k…53 k, Запуск в работу, Available initialisation words – Инструкция по эксплуатации VEGA VEGASON 51K…53K

Страница 62

Advertising
background image

62

VEGASON 51K ... 53K

Запуск в работу

Продолжение плана меню HART ® VEGASON 51 K…53 K

from 1.3.4

1.3.4.8

Available initialisation words:

..SOL

Meas. condition solid

..LIQ

Meas. condition liquid

..FED

Delete false echo memory

..FEN04.58M

False echo e.g. at 4.58 m create new

..FEN48.67FT False echo e.g. at 48.67 ft create new

..FEU03.68M

Extend false echo memory:
add a new false echo to the false echo memory at 3.68

m
..FEU36.05FT add a new false echo to the false echo memory at 36.05
ft

Initialisation word
(enter at last)

newly entered and initialisa-
tion word to enquire with
ENTER and SEND

D

N o t e :
After the adjustment of the initialisation word,
press "ENTER“ and then "SEND“.
Confirm the message to change over to manual
operation with "OK“ and the message to switch
over to automatic operation again with "OK“.
Only then is the adjustment written in the sensor
and effective.

1.4.3.1

1.4.3.1.2

Low:

In case of failure current
output goes to 22 mA

High:

In case of failure current
output goes to < 3,6 mA

1.4.3.1.3

1.4.3.1.3.1

from
1.4.3.

Warning:
Switch off
control!

Generic: SENSOR
Message

..SOL

..FEN04.58M

HE LP D E L E SC E NTE R

Generic: SENSOR
Analogue output

1 A01

13,467mA

2 A0 Alrm typ
3 Loop test
4 D/A trim
5 Scaled D/A trim

HELP SAVE HOME

Generic: SENSOR
Choose analogue output
level

1 4 mA
2 20 mA
3 Other
4 End

Abort Enter

Generic: SENSOR
Fld dev output is
fixed at 4.000 mA

ABORTOK

Generic: SENSOR
A0 Alrm typ
Hold last out valu

Exit