Set receive to transmit delay, Report number of floats, Report receive to transmit delay – Инструкция по эксплуатации VEGA VEGAFLEX 86 Modbus and Levelmaster protocol Coax probe

Страница 87: Коды ошибок

Advertising
background image

87

11 Приложение

VEGAFLEX 86 • Протокол Modbus и Levelmaster

42288-R

U-130320

Set Receive to Transmit Delay

Request:

Параметр

Length

Code/Data

Set Receive to Transmit Delay

7 characters ASCII

UuuRmmm
mmm = milliseconds (50 up to 250),

default = 127 ms

Response:

Параметр

Length

Code/Data

Set Receive to Transmit Delay

6 characters ASCII

UuuROK

Report Number of Floats

Request:

Параметр

Length

Code/Data

Set Receive to Transmit Delay

4 characters ASCII

UuuF

Response:

Параметр

Length

Code/Data

Set Receive to Transmit Delay

5 characters ASCII

UuuFn
n = number of measurement values

(0, 1 or 2)

Report Receive to Transmit Delay

Request:

Параметр

Length

Code/Data

Report Receive to Transmit Delay

4 characters ASCII

UuuR

Response:

Параметр

Length

Code/Data

Report Receive to Transmit Delay

7 characters ASCII

UuuRmmm
mmm = milliseconds (50 up to 250),

default = 127 ms

Коды ошибок

Error Code

Name

EE-Error

Error While Storing Data in EEPROM

FR-Error

Erorr in Frame (too short, too long, wrong data)

LV-Error

Value out of limits

Advertising