Set receive to transmit delay, Report number of floats, Report receive to transmit delay – Инструкция по эксплуатации VEGA VEGAFLEX 81 Modbus and Levelmaster protocol
Страница 93: Коды ошибок
93
11 Приложение
VEGAFLEX 81 • Протокол Modbus и Levelmaster
41828-R
U-130320
Set Receive to Transmit Delay
Request:
Параметр
Length
Code/Data
Set Receive to Transmit Delay
7 characters ASCII
UuuRmmm
mmm = milliseconds (50 up to 250),
default = 127 ms
Response:
Параметр
Length
Code/Data
Set Receive to Transmit Delay
6 characters ASCII
UuuROK
Report Number of Floats
Request:
Параметр
Length
Code/Data
Set Receive to Transmit Delay
4 characters ASCII
UuuF
Response:
Параметр
Length
Code/Data
Set Receive to Transmit Delay
5 characters ASCII
UuuFn
n = number of measurement values
(0, 1 or 2)
Report Receive to Transmit Delay
Request:
Параметр
Length
Code/Data
Report Receive to Transmit Delay
4 characters ASCII
UuuR
Response:
Параметр
Length
Code/Data
Report Receive to Transmit Delay
7 characters ASCII
UuuRmmm
mmm = milliseconds (50 up to 250),
default = 127 ms
Коды ошибок
Error Code
Name
EE-Error
Error While Storing Data in EEPROM
FR-Error
Erorr in Frame (too short, too long, wrong data)
LV-Error
Value out of limits