11 Приложение

POINTRAC 31 • Foundation Fieldbus

41779-R

U-130731

Hex

Quality

Sub-Status

Condition

0x1C

Bad

Out of Service

Transducer block is in mode "Out of Service"
Channel is not assigned to an AIFB
Channel is not available in running application

0x20

Bad

Transducer in MAN

Transducer block is in mode "Manual"

0x44

Uncertain

Last Usable Value

Three or more internal communications have failed

while "Process Data Update". Last usable value is

displayed

0x48

Uncertain

Substitute

On user error codes: 29

0x4C

Uncertain

Initial Value

After startup of device or channel assignment

in AIFB was changed. The channel unit may be

unknown until next "Process Data Update"
On user error codes: 13

0x80

Good (NC)

Non-specific

No errors concerning to channel handling
On user error codes: 0, 33, 45, 71, 126, 127

Mapping of User Error Codes to FF Field Diagnostics

Priority

Description

User Error Codes

NE-107 Status

31

Hardware failure

40, 41

FAILURE

30

Memory failure

34, 35, 36, 37

FAILURE

29

Software failure

80

FAILURE

28

Paramererization corrupt

72

FAILURE

27

Undefined 27

-

-

26

Undefined 26

-

-

25

Paramererization error

16, 17, 25, 52, 53, 57, 66, 117,

120

FAILURE

24

Conflict in MGC

121, 122, 141

FAILURE

23

Communication error in MGC

8

FAILURE

22

MGC slave reports error

38

FAILURE

21

Undefined 21

-

-

20

Undefined 20

-

-

19

Undefined 19

-

-

18

Detector temperature critical

125

OUT_OF_SPEC

17

Error while auto-standardization

73

FAILURE

16

Excessive radiation

123, 124

FAILURE

15

Input out of bounds

13

OUT_OF_SPEC

14

Error while signal processing

68

FAILURE

13

Undefined 13

-

-

12

Undefined 12

-

-

11

Undefined 11

-

-

10

Undefined 10

-

-