Ошибки SCPI

278

FTB-500

–210

“Trigger error”

-----

–211

“Trigger ignored”

Indicates that a GET, *TRG, or triggering signal was
received and recognized by the device but was
ignored because of device timing considerations;
for example, the device was not ready to respond.
Note: a DT0 device always ignores GET and treats
*TRG as a Command Error.

–212

“Arm ignored”

Indicates that an arming signal was received and
recognized by the device but was ignored.

–213

“Init ignored”

Indicates that a request for a measurement initiation
was ignored as another measurement was already
in progress.

–214

“Trigger deadlock”

Indicates that the trigger source for the initiation of a
measurement is set to GET and subsequent
measurement query is received. The measurement
cannot be started until a GET is received, but the
GET would cause an INTERRUPTED error.

–215

“Arm deadlock”

Indicates that the arm source for the initiation of a
measurement is set to GET and subsequent
measurement query is received. The measurement
cannot be started until a GET is received, but the
GET would cause an INTERRUPTED error.

–220

“Parameter error”

Indicates that a program data element related error
occurred. This error message should be used when
the device cannot detect the more specific errors
described for errors –221 through –229.

–221

“Settings conflict”

Indicates that a legal program data element was
parsed but could not be executed due to the current
device state (see IEEE 488.2, 6.4.5.3 and
11.5.1.1.5).

Error

Number

Description

Probable Cause