Справка по командам IEEE 488.2 и специальным командам

244

FTB-500

Обязательные команды IEEE 488.2

*OPC?

Description

The *OPC? query makes possible the
synchronization between the instrument and an
external controller by reading the Output Queue
or by waiting for a service request on the
Message Available (MAV) bit in the Status Byte
Register. The *OPC? query causes the
instrument to place an ASCII character, 1, into its
Output Queue when the device completes all
pending operations. A consequence of this action
is that the MAV bit in the Status Byte Register is
set to state 1.

Syntax

*OPC?

Parameter(s)

None

Response Syntax

<Acknowledge>

Response(s)

Acknowledge:
The response data syntax for <Acknowledge> is
defined as a <NR1 NUMERIC RESPONSE
DATA> element.

The <Acknowledge> response is a single
ASCII-encoded byte corresponding to 1.

The receipt of an <Acknowledge> response
indicates that all pending selected device
operations have been completed.

Example(s)

*OPC? Return 1

See Also

*OPC
*WAI