10 Приложение

VEGASON 62 • Foundation Fieldbus

28791-R

U-130507

CAL_LEVEL_HI

– Min./max.-adjustment: Level at 'Cal_point_hi'. When writing 'Cal_level_hi' and 'Cal_type' = 1

(Online) the 'Cal_point_hi' is automatically set to the current sensor value. The unit is defined in

'Level_unit'

CAL_LEVEL_LO

– Min./max.-adjustment: Level at 'Cal_point_lo'. When writing 'Cal_level_lo' and 'Cal_type' = 1

(Online), the 'Cal_point_lo' is automatically set to the current sensor value. The unit is defined

in 'Level_unit'

CAL_TYPE

– Min./max.-adjustment: Defines type of calibration: Dry: no influence of sensor value. Online:

current sensor value determines 'Cal_point_hi/lo'

level

– Value after min./max. adjustment

level_unit

– Unit code of 'Level', 'Level_offset', 'Cal_level_hi', 'Cal_level_lo'

level_offset

– Offset that is added to the 'Level' value. Unit derives from 'Level_unit'

SENSOR_OFFSET

– Offset that is added to the 'Sensor_value'. Unit derives from 'Sensor_range.unit'

end_of_operation_range

– Set up to suit the process conditions

begin_of_operation_range

– Set up to suit the process conditions

product_type

– Set up to suit the process conditions. If Special-Parameter adjustment has been utilized this

parameter cannot be written

liquids_medium_type

– Set up to suit the process conditions. If Special-Parameter adjustment has been utilized this

parameter cannot be written

solids_medium_type

– Set up to suit the process conditions. If Special-Parameter adjustment has been utilized this

parameter cannot be written

liquids_vessel_type

– Set up to suit the process conditions. If Special-Parameter adjustment has been utilized this

parameter cannot be written

solids_vessel_type

– Set up to suit the process conditions. If Special-Parameter adjustment has been utilized this

parameter cannot be written

fast_level_change

– Set up to suit the process conditions. If Special-Parameter adjustment has been utilized this

parameter cannot be written

first_echo_factor

– Set up to suit the process conditions

pulse_velocity_correction

– Set up to suit the process conditions

echo_quality

– Signal/Noise ratio

empty_vessel_curve_corr_dist

– Distance from the sensor to the product surface. Unit derives from 'Sensor_range.unit'

empty_vessel_curve_corr_op_code

– Update, create new or delete the empty vessel curve

sound_velocity

– Set up to suit the process conditions