Инструкция по эксплуатации Microlife AG1-30

Страница 16

Advertising
background image

10. Technical specifications

Weight: 450g
Size:

175 x 70 x 103mm

Storage temperature:

–20 to +70 °C

85 % relative humidity maximum

Operation temperature:

0 °C to 46 °C

Measuring range:

0 to 300 mmHg

Measuring resolution:

2 mmHg

Accuracy:

within ±3mmHg in 18 °C to 33 °C;

within ±6mmHg in 34 °C to 46 °C

Inflation source:

a volume of at least 200cc to a pressure of

300 mmHg in 4 to 10 sec

Air leakage:

< ± 4mmHg/min

Hysteresis error:

within 0 mmHg to 4 mmHg

Accessories:

1. cuff (adult size with arm circumference of

22 - 32 cm) with inlaid bladder

2. bulb and valve

3. stethoscope (chestpiece attached to the cuff)

4. soft bag

Technical alterations reserved.

13

Advertising