13. Technical Specifications

This device complies with the requirements of the Medical Device
Directive 93/42/EEC.
Technical alterations reserved.

Operating tempera-
ture:

10 - 40 °C / 50 - 104 °F
15 - 95 % relative maximum humidity

Storage temperature: -20 - +55 °C / -4 - +131 °F

15 - 95 % relative maximum humidity

Weight:

560 g (including batteries)

Dimensions:

152 x 92 x 42 mm

Measuring proce-
dure:

oscillometric, corresponding to Korotkoff
method: Phase I systolic, Phase V diastolic

Measurement range: 20 - 280 mmHg – blood pressure

40 - 200 beats per minute – pulse

Cuff pressure display
range:

0 - 299 mmHg

Resolution:

1 mmHg

Static accuracy:

pressure within ± 3 mmHg

Pulse accuracy:

± 5 % of the readout value

Voltage source:

• 4 x 1.5 V Batteries; size AA

• Mains adapter DC 6V, 600 mA (optional)

Reference to
standards:

EN 1060-1 /-3 /-4; IEC 60601-1;
IEC 60601-1-2 (EMC)